Effects of process parameters and geometry on dimensional accuracy and surface quality of thin strut heart valve frames manufactured by laser powder bed fusion

verfasst von: Xiao Zhao, Anqi Liang, Matteo Bellin, Neil W. Bressloff

Erschienen in: The International Journal of Advanced Manufacturing Technology

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Laser powder bed fusion (LPBF) is one of the most popular metal additive manufacturing technologies, which has found its applications in high-value sectors such as aerospace and biomedical devices. Some recent studies on the LPBF of stents have demonstrated its feasibility in the fabrication in thin strut structures including heart valve frames, as used in transcatheter aortic valve implantation (TAVI) for the treatment of severe aortic stenosis. The state of the art method of laser cutting TAVI frame limits the scope for novel concepts which are made possible by additive manufacturing. However, the surface quality and dimensional accuracy of LPBF parts are lower than that produced by laser cutting. To start the development of new TAVI concepts, the feasibility of manufacturing thin frames by LPBF has been investigated based on the SAPIEN 3 frame by Edwards Lifesciences. In this study, simplified frames with strut size from 0.3 to 0.5 mm have been successfully manufactured. The effects of strut size, strut angle, laser power and scan speed on the dimensional accuracy and surface quality were systemically studied. In addition, a representative SAPIEN 3 frame was manufactured and assessed with high-resolution X-ray CT scans. Good overall dimensional accuracy and low porosity were obtained for the SAPIEN 3 frame. However, inclined struts were found to have relatively low dimensional accuracy and poor surface quality.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

1 Introduction

Laser powder bed fusion (LPBF) is one of the most popular metal additive manufacturing technologies. LPBF applies a high-energy laser to melt particles in a thin powder layer to produce three-dimensional parts. Compared to conventional subtractive manufacturing methods, LPBF can produce metallic parts with extremely high structural complexity. Currently, a wide range of metal materials can be processed by LPBF including stainless steel, titanium, titanium alloys, nitinol, aluminium, aluminium alloys and metal matrix composites [13]. LPBF parts typically have high density and fine microstructures, leading to excellent mechanical properties which could be equivalent to that of wrought parts [4]. Thus, LPBF has found its applications in high-value sectors such as aerospace [5, 6] and biomedical devices [79].
Although LPBF has many advantages over the conventional manufacturing methods, there are several major drawbacks, such as large surface roughness [10] and low dimensional accuracy [11], which reduce its popularity in wider applications, especially fabrication of fine-sized products (structure features under 1 mm). Most of the current literature relating to surface roughness are based on large bulk parts. It was reported that the surface finish is directly affected by process parameters, such as laser energy density, building orientation and layer thickness [12]. Khorasani et al. [13] studied the effects of laser power, scan speed, hatch space, scan pattern angle and heat treatment temperature on the top surface roughness of LPBF Ti-6Al-4 V parts via an experiment using the Taguchi method. The results showed that ranking of the most to the least effective parameters was obtained as heat treatment > hatch space > scan speed > laser power > scan pattern angle. Meanwhile, the authors suggested that a combination of higher laser power and lower scan speed produces higher temperatures to reduce the surface tension of melt pool, an effect which could improve the surface quality. Guo et al. [14] investigated the effects of laser power, scan speed and hatch space on surface roughness of as-built IN738LC parts via LPBF. The top surface roughness was obtained between Ra 11 and Ra 28 μm. A minimum surface roughness was obtained with a specific combination of process parameters. The surface became rougher with higher energy density combinations as more mass was transferred to the previously fabricated tracks due to more vigorous Marangoni flow. The surface roughness increased significantly with low energy density combinations since increasing numbers of open pores were formed as well as the presence of many un-melted particles. Snyder and Thole [15] explored the influence of hatch space and contour scan on the roughness of both the up-skin and down-skin surfaces. The results showed the roughness of down-skin surface had higher magnitudes than the up-skin surface (up to 25 μm). They believed that the roughness of up-skin surfaces was tied to the overall size of the melt pool, while the down-skin surfaces were more sensitive to the melt pool depth. Fox et al. [16] reported the roughness of down-skin surfaces increased greatly with smaller overhanging angles. Calignano [17] found downward faces could be successfully built without supports at an angle up to 30° (between surface and horizontal plane) but with a much higher surface roughness and structure deformation.
Dimensional accuracy is another critical issue to consider in the LPBF of fine-sized parts. Tomas et al. [18] investigated the dimensional accuracy of thin-wall, circular H13 tool steel parts (1 mm thickness, diameters of 25–85 mm) with stripe and sectional scan strategies. All the parts were oversized with a dimensional error ranging from 0.1 to 0.5 mm compared to the nominal diameters. Meanwhile, the dimensional error increased along the building height from the base plate. Ahmed et al. [19] fabricated thin-walled AlSi10Mg parts with a thickness from 0.5 to 5 mm. They found both the dimensional error and sample distortion increased with decreasing wall thickness. The sample with design thickness of 0.5 mm had a 2.4% error for actual thickness and 158.62% distortion rate. Charles et al. [20] studied the dimensional error in 45° down-facing surfaces with different process parameters. The study indicated laser power was the most significant factor, followed by layer thickness, scan speed and then the interaction effect of laser power and scan speed. The dimensional error was between 4 and 30%. Wu et al. [21] explored the fabrication limits of thin-wall structures from Ti-6Al-4 V, AlSi10Mg and Inconel 718. The target thickness was set as between 0.1 and 0.5 mm, with inclined angles of 30°, 45° and 60°. The results showed the minimum thicknesses for Ti-6Al-4 V, AlSi10Mg and Inconel 718 were 99 μm, 101 μm and 107 μm, respectively. The maximum inclined angle of 60° can be achieved with specific process parameters.
Generally, for small dimensions which can only be achieved by single-bead mode, the wall thickness is controlled by the melt pool size instead of the CAD design. For larger sizes, achievable by both single-bead mode and raster mode, the final dimensional accuracy is related to the compensation setting of scan strategy and inclined angle as well as the material used.
Since recent research [22, 23] has reported that thin stent structures can be successfully fabricated via LPBF, it raises new opportunities in the fabrication of heart valve frames, which have similar thin strut structures. Transcatheter aortic valve implantation (TAVI) devices—designed for the minimally invasive treatment of a failing aortic valve—typically comprise a stent-like frame and a bioprosthetic valve. The device is commonly delivered into the degenerated valve through femoral access, and then deployed via balloon expansion or self-expansion [24]. State of the art TAVI frames are fabricated by laser cutting from metallic tubes. Figure 1 depicts the balloon-expandable SAPIEN 3 (S3) valve from Edwards Lifesciences. The S3 frame has a 4-layer structure and 12 cells with thin cobalt-chromium struts, approximately 0.3 mm thickness. The inclined struts have an angle of approximately 55° with the vertical direction. The structure provides enough radial strength to push aside the native valve leaflets and can be easily crimped and expanded. The key features of the heart valve frame such as strut thickness and inclined angles are close to the fabrication limits of LPBF, suggesting that improvements are needed in LPBF and/or the geometry of the frame needs to be modified to achieve acceptable quality. However, Melancon et al. [25] reported there are 49% and 41% mismatches between original CAD design and LPBF porous biomaterials for compressive elastic modulus and yield strength due to the large dimensional errors. The capability of LPBF to produce thin strut structure must be fully understood as the baseline in the novel design for heart valve via LPBF.
Recently, we demonstrated that a valve frame manufactured by LPBF could be successfully crimped and expanded [26]. Encouraged by these results, a detailed analysis was conducted on several test models based on the S3 frame. Measurements were made to assess (i) the effect of strut angles and process parameters on the manufacturing quality of these thin strut structures, (ii) the dimensional accuracy and surface morphology of the frames and (iii) the geometric accuracy of a complete S3 style frame using a high-resolution computer tomography (CT) scan.

2 Materials and methods

2.1 Preparation of CAD models

The CAD models used in this study were based on the S3 frame (Edwards Lifesciences). Rhino software (Robert McNeel & Associates, USA) was used to generate the CAD models with a parametric design method via Python scripting. Figure 2a shows a 2D sketch of the S3 frame with its four-layer cell structure. The key features of different layers are illustrated in Fig. 2b–d. More details are listed in Table 1.
Table 1
Key features of the S3 frame
Key features
 
Key features
 
strutWidth (mm)
The width of the struts
strutLowVert (mm)
The length of vertical struts of the 1st layer cell
strutThick (mm)
The thickness of the struts
strutHighVert (mm)
The length of vertical struts of the 4th layer cell
strutAngle
(°)
The angle between inclined strut and vertical direction
slotWidth (mm)
The width of the leaflet assembly slot
intRia (mm)
The radius of small fillet
extRia (mm)
The radius of large fillet
valveRadius
(mm)
The radius of the frame
crownOffset (mm)
The offset height of crown
Two batches of single-layer test models (TMs) were generated based on the 1st layer of the S3 frame, as shown in Fig. 3a, to study the effect of strut angle and process parameters on the quality of the printed thin frames. Meanwhile, a complete model was also constructed with dimensions deemed to be used in the actual S3 frame. The settings of key features of the TM and S3 frames are listed in Table 2.
Table 2
The settings of key features of TM and S3 frames
Features
strutWidth *strutThick (mm)
strutAngle (°)
strutLowVert (mm)
valveRadius = 13 mm
strutHighVert = 4 mm
slotWidth = 1 mm
intRia = 0.1 mm
extRia = 0.2 mm
crownOffset = 0.4 mm
TM1
0.4*0.4
60, 63, 66, 69,72, 75
5
TM2
0.3*0.3, 0.4*0.4, 0.5*0.5
60
5
S3
0.4*0.4
55
2

2.2 Material

Stainless steel (EN 1.4404) was chosen as the printing material due to its low material cost, good biocompatibility and high process maturity. The 316 L SS powder was supplied by LPW Technology Ltd. (Carpenter Additive, US) with the gas atomisation technology. The powder particles have a spherical morphology with a Gaussian size distribution ranging from 10 to 45 μm. Table 3 shows the chemical composition of 316 L SS used in this study.
Table 3
Chemical composition of 316 L SS powder (according to LPW Technology)
Composition
C
Si
Mn
P
S
Cr
Ni
Mo
N
Fe
w.t.%
0.02
0.82
1.33
0.006
0.004
18.2
11.9
2.2
< 0.1
Bal.

2.3 LPBF process

The MySint 100 metal AM system (SISMA S.p.A., Italy) was used for the printing of samples. The system has a cylindrical build volume with a diameter of 100 mm and a height of 100 mm. The laser equipped in the system is a 200 W fibre laser and the laser beam spot diameter is 55 μm. The whole printing process was protected by argon gas with an oxygen concentration below 0.3%. The layer thickness dL was 20 μm. Since the cross-sectional area of struts is very small, there is almost no space for conventional filling vectors as used in the manufacture of large parts. The laser scan vectors shrank with a 75 μm beam compensation, dc, to guarantee the dimensional accuracy of the struts. This effect is discussed in Section 4 with reference to Fig. 13a. Different numbers of concentric borders were used to fill the sliced struts. For 0.3-mm, 0.4-mm and 0.5-mm struts, the numbers of borders were 1, 2 and 4, respectively. The distance between the laser scan vectors was 35 μm. The other process parameters for TM1, TM2 and S3 frames are listed in Table 4. TM1 and S3 frames were fabricated with the same process parameters. TM2 samples were fabricated with different laser powers, P (103–123 W) in Group1 and scan speeds, V (600–800 mm/s) in Group2. Figure 3a shows the test model with supports on the printing platform. Figure 3b illustrates the printed TM2 frames still fixed on the metal build plates.
Table 4
Process parameters for TM and S3 frames
Sample
Laser power (W)
Scan speed (mm/s)
Layer thickness of 20 μm
Border scan without filling scan
TM1
113
700
TM2
Group1
103, 108, 113, 118, 123
700
Group2
113
600, 650, 700, 750, 800
S3
113
700

2.4 Characterisation

The struts of the frames were measured six times with a digital calliper. The average strut width was recorded and compared with the original CAD design. The surface morphology of the frames was examined with the Alicona 3D optical microscope system (Bruker Alicona, Austria). Furthermore, an optical profilometer with a width of 17.519 μm was used to scan the surface of the struts following ISO standard 4288 − 1996. The path length was at least 30 mm to achieve reliable surface roughness data. Finally, the surface roughness was automatically calculated via the surface profiles in the IF-MEASURESUITE software (Bruker Alicona, Austria). The surface roughness of different surfaces was examined for three times with randomly selected struts.
The as-built S3 frame was scanned by a custom Nikon XTEK XTH 225 kVp micro-focus CT system in the µ-VIS lab at the University of Southampton. The beam voltage was set as 200 kV while the current was 115 µA. Three thousand one hundred forty-two projections were taken through 360° with an exposure of 177 ms. The voxel resolution was 16 μm. Segmentation and geometry reconstruction were performed using Avizo Software (Thermo Fisher Scientific, USA) and the exported model was compared with the original CAD model in GOM Inspect 2019 (Carl Zeiss AG, German).

3 Results

3.1 TM1 samples

Figure 4a shows the TM1 frames removed from the build plate. No breakages of struts were observed in the frames. Figure 4b depicts the width of struts of the Batch 1 frames. For the vertical struts, there is minimal effect of strut angle on the manufactured width of the struts, varying between 0.372 ± 0.01 and 0.387 ± 0.012 mm. Thus, the dimensional error of the vertical struts is between 3.3 and 8.8% with a strut angle ranging from 60 to 75°. However, the width of the inclined struts generally increases with strut angle. The width of inclined struts is 0.6 ± 0.043 mm when the strut angle is 60°. The strut width increases to 0.628 ± 0.054 with a strut angle of 63° and fluctuates around 0.75 mm with a strut angle from 66 to 72°. The strut width reaches a maximum of 0.817 ± 0.075 mm with a strut angle of 75°. Compared with the vertical struts, the inclined struts have a much larger dimensional error from 50 to 104% as the strut angle increases from 60 to 75°. Furthermore, the dimensional deviation of the inclined struts is 5–10 times higher than that of the vertical struts. This indicates high manufacturing irregularities resulting from struts at these selected angles.
Figure 5 illustrates the side view of the TM1 frames. The arrow in Fig. 5a shows the build direction. At the down-skin surfaces, some particles are partially melted and sintered to the struts as shown in the red circles. With the increase of strut angle, more partially melted particles and dross are observed on down-skin surfaces as illustrated in Fig. 5b–f. However, the up-skin surfaces are much smoother with few splash drops compared with the down-skin surfaces. What’s more, the inclined struts become more irregular with the increase of strut angle as illustrated in Fig. 5a–f. All the vertical struts show good geometry consistency with a width close to 0.4 mm. However, the irregular surfaces of the inclined struts lead to larger measured widths, which is consistent with the data shown in Fig. 4b. Figure 6a–c depict the 3D surface morphologies of the TM1 frame with 60° strut angle. The colour map shows the variation of the height of the strut surface in the coordinate system of Alicona microscope. As for the up-skin surface, the height value of the surface is generally between 12 and 12.3 mm. The side surface has a height between 27.3 and 27.5 mm, and the down-skin surface has a larger height variation from 10.2 to 10.8 mm. The gradual colour change in Fig. 6a and b could be attributed to the imperfect alignment between strut surface and the optical focus surface. The local colour contrast between valleys and peaks could be used to compare the surface quality of different surfaces. It is clear that the side surface is smoothest with smallest height variation between valleys and peaks. The down-skin surfaces are much rougher compared to the side surface and up-skin surface. The surface roughness was calculated based on the surface profiles and plotted in Fig. 6d. The side surface roughness is not likely to be affected by the strut angle. The side surface roughness fluctuates in a narrow zone between Ra 7.3 and Ra 11.3 μm. However, the up-skin surface roughness increases, varying between Ra 17.7 ± 2.8 and Ra 28.6 ± 5.7 μm when the strut angle increases from 60 to 75°. The down-skin surface roughness is Ra 32.6 ± 11.1 μm with a 60° strut angle. The down-skin surface roughness increases with the strut angle. The surface roughness reaches Ra 58.9 ± 13.5 μm with 72° inclined struts and stays at a high surface roughness of Ra 57.2 ± 10.8 μm with 75° inclined struts. The surface roughness values are summarised in Table 5.
Table 5
Surface roughness of TM1 frames
Strut angle (°)
Surface roughness Ra (µm)
60
63
66
69
72
75
Side surface
8.2 ± 0.06
8.4 ± 1.23
7.3 ± 0.95
9.5 ± 0.81
11.3 ± 2.29
9.4 ± 2.24
Up-skin surface
17.7 ± 2.76
19.7 ± 0.94
21.3 ± 4.76
23.2 ± 3.36
23.8 ± 2.36
28.6 ± 5.66
Down-skin surface
32.6 ± 11.06
35.1 ± 13.85
46.2 ± 14.29
53.1 ± 5.26
58.9 ± 13.45
57.2 ± 10.83

3.2 TM2 samples

Two of the TM2 frames with 0.3 mm struts were observed to have broken cells which are highlighted with red arrows in Fig. 3b. The process parameters used were 600 mm/s and 650 mm/s with fixed laser power of 113 W. All the remaining frames were in good condition. The average measured strut widths of TM2 frames are plotted in Fig. 7. Both laser power and scan speed have a slight effect on the width of vertical struts compared with that on the width of inclined struts. As shown in Fig. 7a and c, vertical struts have good consistent dimensional size with the original CAD design. The dimensional errors are between − 5 and 37 μm with selected laser powers, and similar dimensional errors between − 7 and 35 μm are observed with the variation of scan speed. The width error of vertical struts, barring a few exceptions, slightly increases with laser power but decreases with scan speed. It can be concluded that the 0.3 mm vertical struts have a high dimensional accuracy. However, the dimensional error of vertical struts increases as the TM2 frames became thicker from 0.3 to 0.5 mm. As for the inclined strut, the width increased significantly with laser power except the 0.4-mm and 0.5-mm struts with 108 W. As illustrated in Fig. 7b, the dimensional error of the 0.3 mm inclined struts is between 0.212 and 0.433 mm, namely 71–144% of the designed width. The dimensional error is 36–97% for 0.4-mm frames, while it is 20–61% for 0.5-mm frames. As for the effects of scan speed, the dimensional error decreases at specific scan speeds and then increases, except for the 0.4-mm frames, as shown in Fig. 7d. The dimensional error of inclined struts is 84–126%, 42–77% and 20–44% for 0.3-mm, 0.4-mm and 0.5-mm struts, respectively.
Figure 8 depicts the 3D surface morphologies of three TM2 frames with a laser power of 103 W and a scan speed of 700 mm/s. The large colour differences along the struts observed in Fig. 8a up-skin surface and Fig. 8c down-skin surface indicate an imperfect alignment between strut surface and the optical focus surface. The centre of the struts is generally marginally lower than the edge, which is consistent with the border scan strategy. When the strut thickness is 0.3 mm, a balling effect can be observed at the left side of the strut as shown in the up-skin surface of Fig. 8a. The up-skin surface is generally rougher than side surface but smoother than the down-skin surface. As the strut becomes thicker, all the surfaces generally become smoother as shown in Fig. 8.
Figure 9 plots the variations of surface roughness of the TM2 frames with different laser powers, scan speeds and strut widths. As shown in Fig. 9a, the up-skin surface roughness of 0.3-mm struts gradually decreases with higher laser power than 103 W, but increases with 123 W. The maximum surface roughness is Ra 24.7 ± 3.1 μm, as the minimum is Ra 12.4 ± 0.6 μm. As for 0.4-mm and 0.5-mm struts, the top surface roughness shows similar increase-then-decrease trend with some exceptions. The surface roughness is generally smaller than that of 0.3-mm struts. The minimum surface roughness of Ra 7.7 ± 1.9 μm is achieved with 113 W and 0.5 mm thickness. Compared with the up-skin surfaces, the side surfaces have a smaller surface roughness between Ra 5 and Ra 12 μm. However, there is no evidence that the strut size affects the side surface roughness as illustrated in Fig. 9b. Figure 9c depicts the surface roughness of down-skin surfaces with different laser powers. Generally, the bottom surfaces are much rougher than the top surface and side surface. The surface roughness is between Ra 22.7 ± 1.3 and Ra 53.7 ± 14.1 μm. Similar to the top surface, the bottom surface roughness is prone to be smaller with a thicker strut. The variation of surface roughness with different scan speeds is plotted in Fig. 9d–f. As for the up-skin surface, the surface roughness varies less with faster scan speed compared with the effects of laser power. The side surface roughness is generally within a narrow range except 0.3-mm strut with 700 mm/s and 0.5-mm strut with 650 mm/s. The down-skin surface roughness of 0.3-mm struts increases from Ra 36.3 ± 2.5 to Ra 46.1 ± 11.1 μm as the scan speed increases from 600 to 800 mm/s. However, the bottom surface roughness of 0.4 and 0.5 mm stays between Ra 21 and Ra 27 μm. Although the surface roughness data is very large and complex, some features of the surface quality can be summarised as follows: (a) the side surface is smoothest and the down-skin surface has the largest surface roughness, (2) laser power and scan speed have larger effects on the surface quality of the up-skin and down-skin surfaces than that of the side surface, (3) a thicker strut helps to improve the surface quality.

3.3 S3 frame

Figure 10a depicts the as-built S3 frame removed from the build plate. During manual removal from the plate, there were no strut breakages but the 1st layer of cells was slightly distorted. Figure 10b shows the computational model of the S3 frame, generated from the high-resolution CT scan of the frame, and Fig. 10c depicts the pores in the S3 frame as measured using Avizo software. The CAD model volume of the frame is 94.16 mm3 and the volume of pores is 0.52 mm3. Thus, the porosity of the as-built S3 frame is 0.55%. Most of the pores are found in the inclined struts, crowns and assembly slots, as no pores are found in the vertical struts. Based on the analysis of the pores in Avizo software, there are 3033 pores. From the pore size distribution as shown in Fig. 10d, most of the pores have smaller size than 0.1 mm. The minimum equivalent pore size is 0.018 mm and the maximum pore size is 0.415 mm, which is located in the large sections close to the leaflet assembly slots as shown in Fig. 10c. Meanwhile, most pores were generally found in the centre of the strut cross-sections. This is attributed to the scan strategy used wherein the struts were scanned with two concentric border scans. The CT results suggest that the melt tracks didn’t have sufficient overlap to avoid the creation of pores.
The reconstructed S3 model was exported as a STL file from the Avizo software. Since the STL model is too large to be auto-aligned in GOM Inspect 2019, the initial alignment between the STL file and the designed CAD model was conducted in Rhino. Then, the local best-fit alignment was used in GOM Inspect 2019. The assembly slots were chosen as the target area for alignment. Figure 11a shows the final alignment result. Figure 11b depicts deviation distribution between the printed frame and the design. In both sub-figures, the 1st layer of cells has large deviations due to the distortion of the frame when removing it from the build plate. The distortion also causes deviations of adjacent cells in the 2nd and 3rd layers. Also, the top crowns have larger deviations (0.31 mm) compared to the middle cells (0.04 mm). Figure 11c illustrates the 0.5 mm tolerance distribution on the frame. The red regions highlight where geometric deviation is larger than 0.5 mm. These include the up-skin and down-skin surfaces of the inclined struts, and overhanging surfaces of the leaflet assembly slots, which can be attributed to the large dimensional error described in Sections 3.1 and 3.2.

4 Discussion

Due to the layer-by-layer manufacturing fashion of LPBF, the dimensional accuracy and surface quality of as-built parts are significantly affected by the characteristics of a single melt track and the scan strategy to stack the melt tracks. Since the frame struts are relatively thin, it is difficult to prepare metallurgical samples to observe the cross-section of the struts. Mathematical methods have been used to explore and understand the fundamental behavior of the single melt track in LPBF. The Rosenthal model was first developed in 1941 to evaluate the thermal field during arc welding [27]. Due to the similarities of thermal history between arc welding and LPBF, the Rosenthal model can be used to predict the thermal characteristic in LPBF. There are several assumptions to the model in the simulation of powder-bed simulation [28]: (1) the thermophysical properties are temperature independent. The latent heat due to phase change is neglected. (2) The scanning speed and power input are constant. (3) The heat source is a point source. (4) The heat transfer is only governed by conduction. Convection and radiation are not considered. (5) The deposition of powder has no significant influence on melt pool size. The Rosenthal equation can be expressed as follows:
$$T={T}_{0}+\frac{\epsilon P}{2\pi kr}{exp}^{\left[-\frac{V(r+\xi )}{2\alpha }\right]}$$
(1)
where T0 is the temperature at locations far from the top surface, ε is the absorptivity of the material, P is the laser power, κ is the thermal conductivity, r is the distant from the heat source, and α is the thermal diffusivity. The laser moves along the x-axis, in which the moving coordinate of x-Vt is replaced by ξ. Recently, it has been reported to successfully predict the thermal field, melt track geometry and microstructure fabricated with LPBF [2830]. In this study, a MATLAB program was used to calculate the thermal field of a single melt track based on the Rosenthal model. Table 6 lists the material properties used in the calculation.
Table 6
Material properties used in the calculation
 
Value
Unit
Ref.
Melting point, TM
1385
°C
[31]
Thermal conductivity, κ
12.45
W/mK
[32]
Specific heat, cp
468
J/kgK
[32]
Thermal diffusivity, α
3.4 × 10−6
m2/s
[32]
Absorptivity, ε
0.58
 
[33]
Figure 12a depicts the temperature contour and melt track boundary of a single melt track using 113 W and 700 mm/s, obtained using the MATLAB calculation from The MathWorks, Inc. (USA). The width of the melt track is 131.4 μm, while the depth is 65.7 μm. As shown in Fig. 12b, the width and depth of the melt track increase linearly with the increase of laser power. The width and depth of the melt track vary, respectively, between 125.3 and 137.2 μm, and 62.6 and 68.6 μm. With respect to variation of the scan speed between 600 and 800 mm/s (Fig. 12c), the width of the melt track decreases linearly from 141.7 to 123 μm whilst the depth decreases from 70.8 to 61.5 μm.
Figure 13a and b schematically illustrate the border scan strategy for vertical struts and inclined struts. The cross-sections of struts are simplified as square and rectangle shapes due to the large ratio between the strut width and radius of strut curvature. As shown in the top view, the dash lines represent the border of the struts. The vertical struts have a square shape since the width and thickness of struts were set as the same value. The longer side of the cross-section of inclined struts can be calculated as strutWidth/cos(strutAngle), roughly 1.22–3.87 times strutWidth as the strutAngle increases from 55 to 75°. The cyan solid lines represent the laser scan vectors, and the yellow solid shape is the melt track of the border. The red double arrow represents the laser beam compensation, dc. As shown in the side view of struts, the melt tracks are stacked with the specific layer thickness, 20 μm. As for the inclined struts, the melt tracks shift in every single layer with an offset of dL*tan(strutAngle), roughly 28–75 μm.
Due to the laser beam energy distribution in LPBF, the cross-section of melt tracks is generally symmetric [34]. Figure 13c depicts the cross-sectional view of the melt tracks in the 0.3-mm strut. Based on the calculated results in Fig. 12, the depth of the melt track is much higher than the layer thickness. The melt track tends to re-melt at least three previous layers. The middle of the border melt track has a compensation, dc, with the geometry border (dash line). For the vertical struts, the width of vertical strut can be estimated as strutWidth-2dc+ width. Based on the calculated results, the width of vertical strut fabricated with 113 W and 700 mm/s is 0.281 mm. The experimental results in TM2 samples are 0.298 mm and 0.302 mm, which are a little higher than the calculated results. The effects of laser power and scan speed on the vertical strut width are consistent with the calculated results from the Matlab. As for the inclined struts, the slice shifts layer-by-layer as discussed in the previous paragraph. Part of the scan vectors are directly on the powder particles. Compared with the vertical strut, less solid beneath the melt track leads to lower heat conduction [17] and higher sustained temperatures in the melt pool. Thus, more powder particles will be melted and form a larger melt track. The excess melted mass tends to penetrate into the powder particles with the effect of gravity and surface capillary force [35], which forms large dross as shown in Figs. 5b–f and 13c. Meanwhile, the depth of the melt track also causes the severe mismatch of melt tracks and the designed geometry. With the increase of laser power, the melt track becomes wider and deeper, which leads to a larger dimensional error for the inclined struts. However, the effects of scan speed on the strut width does not follow the MatLab calculated results. It was found that increasing scan speed gave rise to residual stresses [36], which can worsen the warping effect [37] and cause larger dimensional errors. Furthermore, for the smaller planar cross-section of the vertical strut, the overlaps between melt tracks are higher than the inclined struts and the large cross-section near the assembly slots. When the struts become thicker, the overlaps between melt track become smaller, which could lead to pores being created in the frames. This helps to explain the specific distribution of pores in the S3 frame as illustrated in Fig. 10c. Also, thicker struts have better structural stability when interacting with the powder spreader. Thus, the dimensional accuracy is generally better with thicker struts.
Unlike conventional manufacturing technologies, LPBF tends to produce surfaces with several topographical features including micro-scaled surface asperity, wave-like melt track features and form error globules of surface pores [12]. The surface profiles of the side surface, up-skin surface and down-skin surface are also marked in Fig. 13c. The wave-like features and form error globules are the main factors to influence the surface roughness in this study. There is minimum variation in the waviness profile of the side surface, but it is higher in the up-skin surface and higher still in the down-skin surface. There are less partially sintered particles on the side surface and up-skin surface due to the lower sustained melt pool temperature. However, many partially melted particles and even large dross can be observed on the down-skin surface, which is also reported in other publications [11, 12, 15, 35]. Thus, the bottom surface is much rougher compared to the other surfaces. When strut thickness increases, there is improved structural stability and better thermal conduction through the struts helps to reduce the surface roughness. The effects of laser power and scan speed variation on the surface roughness is unclear based on our results. However, some basic features of surface quality of thin strut structures have been concluded in Section 3.2.

5 Conclusion

In this study, the effects of strut dimensions and angles, laser power and scan speed on the dimensional accuracy and surface quality of the thin struts, that make up a typical frame for a replacement heart valve, were systemically studied. The results can be concluded as follows:
1)
Thin strut heart valve frames with strut cross-sectional dimensions between 0.3 and 0.5 mm can be successfully manufactured using laser powder bed fusion, LPBF.
 
2)
Relative to laser power, scan speed and cross-sectional dimensions, the angle of inclined struts has the largest impact on dimensional accuracy and good surface quality. Although a strut angle of 75° can be achieved by LPBF, a much smaller strut angle is recommended for frames to be manufactured using LPBF.
 
3)
Vertical struts have better dimensional accuracy and surface quality than inclined struts. The down-skin surfaces have the largest surface roughness, in excess of Ra 20 μm. All frames analysed here have not been treated using sand blasting, polishing or any other post process. It is clear that surface treatment will be necessary to achieve the high surface quality needed for replacement heart valves.
 
4)
The complete S3 frame had a low porosity of 0.55% and a generally good dimensional accuracy when compared to the original computer model. However, improvements are required for inclined struts and overhanging surfaces such as those found in the leaflet assembly slots.
 

Acknowledgements

The authors acknowledge the µ-VIS X-ray Imaging Centre (supported by EPSRC grant EP/H01506X/1) and the Biomedical Imaging Unit at the University of Southampton, for the provision of imaging, data processing and management infrastructure and expertise.

Declarations

Competing interests

Xiao Zhao, Anqi Liang and Neil W. Bressloff have no relevant financial or non-financial interests to disclose. Matteo Bellin is employed by SISMA S.p.A.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
10.
Zurück zum Zitat Kruth JP, Yasa E, Deckers J (2008) Roughness improvement in selective laser melting. Proceedings of the 3rd international conference on polymers and moulds innovations, 170–183 Kruth JP, Yasa E, Deckers J (2008) Roughness improvement in selective laser melting. Proceedings of the 3rd international conference on polymers and moulds innovations, 170–183
24.
Zurück zum Zitat Giordano A, Biondi-Zoccai G, Frati G (2019) Transcatheter aortic valve implantation: clinical, interventional and surgical perspectives. Springer, Switzerland Giordano A, Biondi-Zoccai G, Frati G (2019) Transcatheter aortic valve implantation: clinical, interventional and surgical perspectives. Springer, Switzerland
25.
27.
Zurück zum Zitat Rosenthal D (1941) Mathematical theory of heat distribution during welding and cutting. Weld J 20:220–234 Rosenthal D (1941) Mathematical theory of heat distribution during welding and cutting. Weld J 20:220–234
30.
Zurück zum Zitat Montgomery C, Beuth J, Sheridan L, Klingbeil N Process mapping of Inconel 625 in laser powder bed additive manufacturing. In2015 International Solid Freeform Fabrication Symposium 2015. University of Texas at Austin Montgomery C, Beuth J, Sheridan L, Klingbeil N Process mapping of Inconel 625 in laser powder bed additive manufacturing. In2015 International Solid Freeform Fabrication Symposium 2015. University of Texas at Austin
31.
Zurück zum Zitat Mills KC (2002) Recommended values of thermophysical properties for selected commercial alloys. Woodhead Publishing, CambridgeCrossRef Mills KC (2002) Recommended values of thermophysical properties for selected commercial alloys. Woodhead Publishing, CambridgeCrossRef
Metadaten
Titel
Effects of process parameters and geometry on dimensional accuracy and surface quality of thin strut heart valve frames manufactured by laser powder bed fusion
verfasst von
Xiao Zhao
Anqi Liang
Matteo Bellin
Neil W. Bressloff
Publikationsdatum
16.05.2024
Verlag
Springer London
Erschienen in
The International Journal of Advanced Manufacturing Technology
Print ISSN: 0268-3768
Elektronische ISSN: 1433-3015
DOI
https://doi.org/10.1007/s00170-024-13627-3

    Marktübersichten

    Die im Laufe eines Jahres in der „adhäsion“ veröffentlichten Marktübersichten helfen Anwendern verschiedenster Branchen, sich einen gezielten Überblick über Lieferantenangebote zu verschaffen.